Sigma Spectrum Infusion System FirmwareOperating system · Baxter

CVE-2014-5431

MEDIUM · 6.8 CVSS v3.0 Published 2019-03-26
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
73/100
Remediation priority · Elevated
No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 contains a hard-coded password, which provides access to basic biomedical information, limited device settings, and network configuration of the WBM, if connected. The hard-coded password may allow an attacker with physical access to the device to access management functions to make unauthorized configuration changes to biomedical settings such as turn on and off wireless connections and the phase-complete audible alarm that indicates the end of an infusion phase. Baxter has released a new version of the SIGMA Spectrum Infusion System, version 8, which incorporates hardware and software changes.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

The Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module version 16 contains a hard-coded password that provides unauthorized access to basic biomedical information, limited device settings, and network configuration of the WBM. An attacker with physical access to the device can use this credentials to access management functions and make unauthorized changes to biomedical settings including toggling wireless connections and the phase-complete audible alarm.

MitigationUpgrade to version 8 of the SIGMA Spectrum Infusion System which incorporates hardware and software changes to address the hard-coded password vulnerability. Until upgrade is possible, implement strict physical access controls and restrict network connectivity to the device.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Sigma Spectrum Infusion System FirmwareOperating system
Affected:= 6.05

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Physical
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify the device model
    Locate the model number on the Baxter SIGMA Spectrum Infusion System physical label or device documentation. Look for model number 35700BAX.
    Affected if The device is a Baxter SIGMA Spectrum model 35700BAX.
  2. Check the installed firmware version
    Access the device's system information menu or boot screen to view the firmware version. On the SIGMA Spectrum, this is typically found under Settings > System Info or during the device startup sequence.
    Affected if The firmware version is exactly 6.05.
  3. Verify the wireless battery module version
    Access the wireless battery module (WBM) information through the device's settings menu or diagnostics screen. The WBM version is typically displayed in the network or wireless configuration section.
    Affected if The wireless battery module version is 16.
  4. Confirm physical and network access posture
    Review physical access controls and network segmentation surrounding the infusion pump. Document whether the device is accessible to unauthorized personnel or connected to unrestricted networks.
    Affected if The device has unrestricted physical access or is connected to untrusted networks, allowing potential exploitation of the hard-coded credential.

The environment is affected if the device is a Baxter SIGMA Spectrum model 35700BAX running firmware version 6.05 with wireless battery module version 16, and the device lacks adequate physical access controls.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Upgrade to version 8 of the SIGMA Spectrum Infusion System which incorporates hardware and software changes to address the hard-coded password vulnerability. Until upgrade is possible, implement strict physical access controls and restrict network connectivity to the device.

Fix this in Sigma Spectrum Infusion System Firmware Scoped from the published advisory
  • Consultation12.0 h
  • Implementation20.0 h
  • Testing16.0 h
  • Review / QA8.0 h
56.0 hours of engineering $9,840
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $15,744.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2014-5431 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2014-5431 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data