Plum A\+ Infusion System FirmwareOperating system · Pifzer

CVE-2015-3953

CRITICAL · 9.8 CVSS v3.0 Published 2019-03-25
Fix available
A fix is available. Upgrade to after 13.6 or later.
See remediation →
100/100
Remediation priority · Urgent
Remotely reachable No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Hard-coded accounts may be used to access Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

Hard-coded authentication credentials exist in Hospira Plum A+, Plum A+3, and Symbiq infusion pump systems, allowing unauthorized access via FTP (port 20) and Telnet (port 23). The credentials are embedded in the firmware and cannot be changed by users.

MitigationBlock or close Port 20/FTP and Port 23/TELNET at the network perimeter to prevent external access; consider migrating to the Plum 360 Infusion System which is not affected.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Plum A\+ Infusion System FirmwareOperating system
Affected:<= 13.4
Plum A\+3 Infusion System FirmwareOperating system
Affected:<= 13.6
Symbiq Infusion System FirmwareOperating system
Affected:<= 3.13

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Network
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify the infusion pump model
    Locate the device label or access the device's system information to confirm the model is a Hospira/Pfizer Plum A+, Plum A+3, or Symbiq Infusion System
    Affected if The device is one of these three models
  2. Check the installed firmware version
    Access the device's administrative interface or firmware settings to determine the firmware version number
    Affected if Firmware version is 13.4 or lower for Plum A+, 13.6 or lower for Plum A+3, or 13.3 or lower for Symbiq
  3. Verify if FTP service is exposed
    Scan the network for open port 20 or conduct a port scan against the device to check if FTP is listening
    Affected if Port 20/FTP is open and accessible on the device
  4. Verify if Telnet service is exposed
    Scan the network for open port 23 or conduct a port scan against the device to check if Telnet is listening
    Affected if Port 23/Telnet is open and accessible on the device

You are affected if you have a Plum A+, Plum A+3, or Symbiq infusion pump running the specified firmware versions AND FTP (port 20) or Telnet (port 23) services are accessible on your network.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

dbcve · scoped
Upgrade available Upgrade to a release after 13.6
Interim mitigation

Block or close Port 20/FTP and Port 23/TELNET at the network perimeter to prevent external access; consider migrating to the Plum 360 Infusion System which is not affected.

Recommended fix High confidence
  1. Close Port 20/FTP on the affected Plum A+, Plum A+3, and Symbiq Infusion Systems to prevent unauthorized access via the hard-coded accounts.
  2. Close Port 23/TELNET on the affected systems to block remote access using the hard-coded credentials.
  3. Verify that the port closures are properly applied and that remote access is no longer possible on these ports.
  4. Consider migrating to the Plum 360 Infusion System, which Hospira states is not vulnerable to this issue.
  5. Implement network segmentation to isolate infusion devices from untrusted networks.
  6. Change any default credentials if identified through a thorough security assessment.
Caveat No direct firmware upgrade is available for the affected Plum A+, Plum A+3, or Symbiq Infusion Systems; the vendor recommends migrating to Plum 360 or applying network-based mitigations

Generated from the published advisory — verify against the referenced sources before acting.

Fix this in Plum A\+ Infusion System Firmware Scoped from the published advisory
  • Consultation8.0 h
  • Implementation4.0 h
  • Testing4.0 h
  • Review / QA2.0 h
18.0 hours of engineering $3,280
Get the upgrade done

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $5,248.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2015-3953 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2015-3953 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data