Medfusion 4000 Wireless Syringe Infusion PumpOperating system · Smiths Medical

CVE-2017-12726

HIGH · 7.3 CVSS v3.0 Published 2018-02-15
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
82/100
Remediation priority · High
Remotely reachable No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

The Medfusion 4000 Wireless Syringe Infusion Pump versions 1.1, 1.5, and 1.6 contain hardcoded credentials for the built-in Telnet service. An attacker with network access to the pump can authenticate using these hardcoded credentials to access the communications module, potentially intercepting or manipulating infusion pump communications.

MitigationRestrict network access to the infusion pump using network segmentation and firewalls; disable Telnet if external communications are not required; contact Smiths Medical for firmware updates that address the hardcoded credentials.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Medfusion 4000 Wireless Syringe Infusion PumpOperating system
Affected:= 1.1= 1.5= 1.6

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Network
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
Low
Integrity
Low
Availability
Low

CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify the device model
    Locate the infusion pump and check the physical label or device display for the model name. Confirm it is a Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump.
    Affected if The device is a Medfusion 4000 pump
  2. Check the installed firmware version
    Access the pump's system information menu or check the device label/packaging for the firmware version number. Compare it to the affected versions: 1.1, 1.5, and 1.6.
    Affected if The firmware version is 1.1, 1.5, or 1.6
  3. Determine if Telnet service is enabled
    Check the pump's network or communications settings menu for Telnet service status. If accessible via the administrative interface, look for Telnet or remote access configuration options.
    Affected if Telnet service is enabled on the pump
  4. Verify network accessibility of the Telnet port
    Attempt a TCP connection to port 23 on the pump's IP address from an external system, or review network firewall rules to determine if inbound connections to port 23 are permitted toward the pump.
    Affected if Port 23 (Telnet) is reachable from a network where an attacker could reside

The environment is affected if you have a Medfusion 4000 pump running firmware version 1.1, 1.5, or 1.6 with Telnet enabled and network-accessible on port 23.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Restrict network access to the infusion pump using network segmentation and firewalls; disable Telnet if external communications are not required; contact Smiths Medical for firmware updates that address the hardcoded credentials.

Fix this in Medfusion 4000 Wireless Syringe Infusion Pump Scoped from the published advisory
  • Consultation4.0 h
  • Implementation8.0 h
  • Testing4.0 h
  • Review / QA2.0 h
18.0 hours of engineering $3,200
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $5,120.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2017-12726 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2017-12726 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data