Intellivue Mx40 FirmwareOperating system · Philips

CVE-2017-9658

MEDIUM · 6.5 CVSS v3.0 Published 2018-04-30
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
70/100
Remediation priority · Elevated
No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Certain 802.11 network management messages have been determined to invoke wireless access point blacklisting security defenses when not required, which can necessitate intervention by hospital staff to reset the device and reestablish a network connection to the Wi-Fi access point. During this state, the Philips IntelliVue MX40 Version B.06.18 can either connect to an alternative access point within signal range for association to a central monitoring station, or it can remain in local monitoring mode until the device is reset by hospital staff. CVSS v3 base score: 6.5, CVSS vector string: AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H. Philips has released software update, Version B.06.18, to fix the improper cleanup on thrown exception vulnerability, and implement mitigations to reduce the risk associated with the improper handling of exceptional conditions vulnerability. The software update implements messaging and alarming on the MX40 and at the central monitoring station, when the MX40 disconnects from the access point.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

The Philips IntelliVue MX40 Version B.06.18 improperly handles certain 802.11 network management messages, causing unnecessary wireless access point blacklisting. This triggers a denial-of-service condition where the device disconnects and requires manual intervention by hospital staff to reset, impacting availability of patient monitoring during the disconnection period.

MitigationApply Philips software update Version B.06.18 to the MX40 devices, which implements proper exception handling and adds messaging/alarming when the device disconnects from the access point.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Intellivue Mx40 FirmwareOperating system
Affected:< b.06.18

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Adjacent
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
None
Availability
High

CVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify the device model
    Locate the IntelliVue MX40 patient monitor unit and verify its model number via the device label, clinical system inventory, or network discovery
    Affected if The device is not a Philips IntelliVue MX40 monitor
  2. Determine the firmware version
    Access the device's system information menu, clinical workstation interface, or device management software to retrieve the installed firmware version
    Affected if The firmware version is B.06.18 or earlier (versions below B.06.18)
  3. Check for manual reset incidents
    Review maintenance logs, incident reports, or nursing documentation for entries indicating the MX40 required manual reset or reconnection by hospital staff
    Affected if The device has required manual intervention to restore connectivity after unexpected disconnections
  4. Verify wireless connectivity issues
    Examine network logs or wireless controller alerts for repeated disconnections or access point blacklisting events involving the MX40 device
    Affected if The device shows repeated unexplained disconnections from wireless access points

A user is affected if they have an IntelliVue MX40 device running firmware version B.06.18 or earlier that has experienced unexplained wireless disconnections requiring manual reset.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Apply Philips software update Version B.06.18 to the MX40 devices, which implements proper exception handling and adds messaging/alarming when the device disconnects from the access point.

Fix this in Intellivue Mx40 Firmware Scoped from the published advisory
  • Consultation4.0 h
  • Implementation2.0 h
  • Testing6.0 h
  • Review / QA3.0 h
15.0 hours of engineering $2,600
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $4,160.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2017-9658 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2017-9658 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data