Intellivue Mp2 FirmwareOperating system · Philips

CVE-2018-10601

HIGH · 8.2 CVSS v3.1 Published 2018-06-05
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
87/100
Remediation priority · High
No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
IntelliVue Patient Monitors MP Series (including MP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only), and Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3 have a vulnerability that exposes an "echo" service, in which an attacker-sent buffer to an attacker-chosen device address within the same subnet is copied to the stack with no boundary checks, hence resulting in stack overflow.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

Stack-based buffer overflow in the exposed 'echo' service of Philips IntelliVue patient monitors (MP Series, MX Series, and Avalon FM monitors). An attacker on the same subnet can send a specially crafted buffer to any affected device that gets copied to the stack without boundary checking, leading to potential code execution.

MitigationIsolate affected medical devices on a dedicated network segment with strict access controls; restrict network connectivity to only essential systems and apply vendor patches when available.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Intellivue Mp2 FirmwareOperating system
Affected:all versions
Intellivue X2 FirmwareOperating system
Affected:all versions
Intellivue Mp30 FirmwareOperating system
Affected:all versions
Intellivue Mp50 FirmwareOperating system
Affected:all versions
Intellivue Mp70 FirmwareOperating system
Affected:all versions
Intellivue Np90 FirmwareOperating system
Affected:all versions
Intellivue Mx700 FirmwareOperating system
Affected:all versions
Intellivue Mx800 FirmwareOperating system
Affected:all versions

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Adjacent
Complexity
High
Privileges
None
User interaction
None
Scope
Changed
Confidentiality
High
Integrity
Low
Availability
High

CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify the device model
    Check the device label, web interface, or query the device via SNMP (OID 1.3.6.1.4.1.341.7.1.1.2.0 for model name) to confirm it is one of: MP2, X2, MP30, MP50, MP70, NP90, MX700, or MX800
    Affected if Device model matches one of the listed IntelliVue monitors
  2. Verify the echo service is exposed
    Use network scanning tool (e.g., nmap) to probe the device for open ports. The echo service typically listens on port 7 (TCP/UDP). Run: nmap -p 7 -sV <device-ip>
    Affected if Port 7 (echo) is open and responding on the target device
  3. Confirm network accessibility
    Determine if the device is reachable from your network segment. Use ping or nmap to verify the device IP is accessible. Check if the device shares a subnet with untrusted systems
    Affected if Device is reachable on the network and could receive packets from an attacker on the same subnet
  4. Check firmware version (optional)
    If accessible via web interface or SNMP, query the firmware version (OID 1.3.6.1.4.1.341.7.1.1.6.0). Note: all firmware versions are affected per the advisory
    Affected if Device is any firmware version of the affected models

You are affected if you have a Philips IntelliVue MP2, X2, MP30, MP50, MP70, NP90, MX700, or MX800 monitor with the echo service (port 7) exposed on a network accessible to potential attackers.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Isolate affected medical devices on a dedicated network segment with strict access controls; restrict network connectivity to only essential systems and apply vendor patches when available.

Fix this in Intellivue Mp2 Firmware Scoped from the published advisory
  • Consultation12.0 h
  • Implementation24.0 h
  • Testing16.0 h
  • Review / QA8.0 h
60.0 hours of engineering $10,560
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $16,896.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2018-10601 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2018-10601 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data