CVE-2019-25719
Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.
NVD · uneditedDräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.
dbcve analysis · high confidenceNetwork message handling vulnerabilities in Dräger Infinity patient monitors (VG4.1.1 and lower) allow network-adjacent attackers to spoof or tamper with data, modify alarm settings/limits, and cause denial-of-service by overwhelming the device with traffic, leading to reboot and loss of network functionality.
Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.
CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.
From the vector- Attack vector
- Network
- Complexity
- Low
- Privileges
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- Low
- Integrity
- Low
- Availability
- High
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.
dbcve checksWork through these to decide whether this CVE applies to you.
-
Identify Dräger Infinity monitor model and firmware versionAccess the device's system information menu or service interface to retrieve the exact firmware version number. Compare against VG4.1.1 - versions at or below this threshold are affected.Affected if Firmware version is VG4.1.1 or lower
-
Verify Infinity network ports are enabledCheck device network configuration settings to confirm if Infinity protocol ports (typically UDP/TCP ports used for patient monitor networking) are in an enabled state.Affected if Infinity network ports are enabled and accessible
-
Confirm wireless network functionality is activeInspect wireless configuration to determine if the monitor has an active wireless adapter connected to any AP.Affected if Wireless connectivity is enabled and the device is within radio range of accessible APs
-
Assess network exposure of the deviceReview network infrastructure configuration to determine if the monitor is on a shared clinical network segment accessible from adjacent network locations.Affected if Device is on a network segment accessible to unauthorized adjacent attackers
A user is affected if their Dräger Infinity monitor runs firmware VG4.1.1 or lower with Infinity network ports or wireless connectivity enabled and accessible to adjacent attackers.
Generated from the published advisory. Verify against your own configuration.
Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.
From vendor dataIsolate affected devices on dedicated network segments, disable unused Infinity network ports, implement network monitoring for anomalous traffic patterns, and coordinate with Dräger for available software updates or vendor-specific hardening guidance.
- Consultation6.0 h
- Implementation12.0 h
- Testing6.0 h
- Review / QA3.0 h
An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $7,680.
Scan for this in your stack
Free · runs locallyCheck whether your project pulls in CVE-2019-25719 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.
References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.
Primary sourcesPractitioner notes
ContributedPeer-ranked notes from engineers who’ve handled CVE-2019-25719 in production — separate from our analysis above.
The advisory tells you what broke. It rarely tells you what actually worked. If you’ve dealt with this one, that detail is what the next engineer is searching for.
- The version that genuinely resolved it — not the one the vendor claimed
- A config change or rule that shut the vector down
- A gotcha in the upgrade path that cost you an afternoon
No notes yet
Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.
A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.
- Verified mitigations, workarounds, and config changes
- Version or environment caveats, and links to real fixes
- No weaponised exploit code, or anything meant to cause harm
- No spam, self-promotion, credentials, or personal data