Uncontrolled Resource ConsumptionWeakness · CWE-400

CVE-2019-25724

MEDIUM · 6.5 CVSS v3.1 Published 2026-06-02
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
70/100
Remediation priority · Elevated
No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Dräger Infinity M300 patient worn monitors with software version VG2.x and earlier contain a network-based denial of service vulnerability that allows attackers with access to the hospital or Infinity Network to repeatedly trigger device reboots until the device enters a fail state requiring manual restart. Attackers can exploit this vulnerability to cause loss of wireless network connectivity, temporary loss of patient monitoring, and interruption of alarm functionality until the device is manually recovered.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · moderate confidence

The Dräger Infinity M300 patient monitor contains a network-based denial of service vulnerability where attackers with access to the hospital or Infinity Network can repeatedly trigger device reboots, eventually causing the device to enter a fail state that requires manual restart. This leads to loss of wireless connectivity, interruption of patient monitoring, and loss of alarm functionality.

MitigationImplement network segmentation to isolate medical devices from untrusted networks and apply strict access controls to prevent unauthorized network access to the Infinity Network. Monitor for abnormal reboot patterns and consider firmware updates if available from Dräger.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Adjacent
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
None
Availability
High

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify the device model
    Locate the physical device or network inventory to confirm it is a Dräger Infinity M300 patient monitor. Check device labels, documentation, or network scanning results for the exact model identifier.
    Affected if The device is an Infinity M300 monitor and is present on your network.
  2. Determine the software version
    Access the device's system information menu or management interface to identify the installed VG version number. Compare this version to the affected range of VG2.x and earlier.
    Affected if The installed VG version is VG2.x or any version prior to VG2 (e.g., VG1.x).
  3. Verify network accessibility
    Check network configuration settings on the device or corresponding network infrastructure to determine if the device has an IP address and is reachable on the hospital or Infinity Network.
    Affected if The device is connected to a network without isolation from untrusted or unauthorized network segments.
  4. Inspect network segmentation
    Review network infrastructure configuration (VLANs, firewalls, access control lists) to determine if the device is isolated from non-medical or untrusted network segments.
    Affected if The device shares a network segment with general hospital systems or is accessible from networks outside the intended Infinity Network.
  5. Check for reboot patterns
    Review device event logs, network monitoring logs, or system uptime records for evidence of repeated or unexplained reboots occurring in short succession.
    Affected if Logs show frequent unexpected reboots or device fail states requiring manual restart.

A user is affected if they have a Dräger Infinity M300 monitor running software version VG2.x or earlier that is accessible on a network segment without proper isolation from untrusted access.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Implement network segmentation to isolate medical devices from untrusted networks and apply strict access controls to prevent unauthorized network access to the Infinity Network. Monitor for abnormal reboot patterns and consider firmware updates if available from Dräger.

Have this fixed Scoped from the published advisory
  • Consultation8.0 h
  • Implementation16.0 h
  • Testing12.0 h
  • Review / QA6.0 h
42.0 hours of engineering $7,360
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $11,776.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2019-25724 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2019-25724 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data