Iks G6824a FirmwareOperating system · Moxa

CVE-2019-6520

HIGH · 7.5 CVSS v3.1 Published 2019-03-05
Fix available
A fix is available. Upgrade to after 4.5 or later.
See remediation →
84/100
Remediation priority · High
Remotely reachable No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Moxa IKS and EDS does not properly check authority on server side, which results in a read-only user being able to perform arbitrary configuration changes.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · moderate confidence

Moxa IKS and EDS industrial Ethernet switches contain a server-side improper authorization vulnerability where the server fails to properly enforce permission boundaries. A read-only user account can bypass authorization checks to perform administrative-level configuration changes, effectively achieving privilege escalation through the web interface or management interface.

MitigationApply vendor-supplied firmware patches when available; until then, restrict management access to trusted users only and consider network segmentation to limit exposure of the management interface.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Iks G6824a FirmwareOperating system
Affected:<= 4.5
Eds 405a FirmwareOperating system
Affected:<= 3.8
Eds 408a FirmwareOperating system
Affected:<= 3.8
Eds 510a FirmwareOperating system
Affected:<= 3.8

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Network
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
None
Integrity
High
Availability
None

CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify device model and firmware version
    Access the device web interface or management console and locate the firmware version information, typically found under System > Device Information or a similar status/overview page
    Affected if The firmware version matches one of the affected products: IKS G6824a with firmware <= 4.5, or EDS 405a/408a/510a with firmware <= 3.8
  2. Check for presence of read-only user accounts
    Navigate to the user management or account settings section in the web interface and review the list of defined user accounts and their assigned privilege levels
    Affected if One or more read-only or low-privilege user accounts exist on the device, as these accounts can exploit the authorization bypass
  3. Verify management interface accessibility
    Determine if the web-based management interface is reachable from the network by attempting to access the device IP via HTTP/HTTPS or checking firewall rules and network access controls
    Affected if The management interface is exposed to untrusted networks, as the vulnerability is exploited through this interface
  4. Audit configuration change logs for privilege escalation
    Review system logs, audit logs, or configuration change logs to identify any administrative-level configuration changes made by accounts with read-only or limited privileges
    Affected if Configuration changes were performed by accounts that should only have read-only permissions, indicating successful exploitation of the authorization bypass

A user is affected if they operate a Moxa IKS G6824a or EDS 405a/408a/510a device with firmware at or below the affected version AND have read-only user accounts configured with access to the management interface.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Upgrade available Upgrade to a release after 4.5
Interim mitigation

Apply vendor-supplied firmware patches when available; until then, restrict management access to trusted users only and consider network segmentation to limit exposure of the management interface.

Fix this in Iks G6824a Firmware Scoped from the published advisory
  • Consultation2.0 h
  • Implementation4.0 h
  • Testing3.0 h
  • Review / QA2.0 h
11.0 hours of engineering $1,930
Get the upgrade done

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $3,088.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2019-6520 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2019-6520 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data