Zoom Latitude Pogrammer\/recorder\/monitor 3120 FirmwareOperating system · Bostonscientific

CVE-2021-38400

MEDIUM · 6.8 CVSS v3.1 Published 2021-10-04
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
73/100
Remediation priority · Elevated
No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
An attacker with physical access to Boston Scientific Zoom Latitude Model 3120 can remove the hard disk drive or create a specially crafted USB to extract the password hash for brute force reverse engineering of the system password.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · moderate confidence

The Boston Scientific Zoom Latitude Model 3120 (a cardiac device programmer) stores system passwords in a way that allows extraction from the hard disk drive via physical access. An attacker with physical access can remove the HDD or use a crafted USB device to obtain password hashes, which can then be brute-forced to recover the system password and gain unauthorized access to the device and its clinical functions.

MitigationPhysical access controls (restricted access to device storage, device locking) are the primary mitigations. Contact Boston Scientific for potential firmware updates or device replacement options, as this is a legacy device that may have limited vendor support.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Zoom Latitude Pogrammer\/recorder\/monitor 3120 FirmwareOperating system
Affected:all versions

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Physical
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify Zoom Latitude Model 3120 devices in your environment
    Inventory all medical devices and clinical workstations. Confirm the presence of Boston Scientific Zoom Latitude Programmer/Recorder/Monitor model 3120 devices. Check device labels, asset inventories, or clinical equipment databases for this specific model number.
    Affected if The device model 3120 is present in your facility
  2. Verify physical access controls to the device
    Inspect the physical location of the device. Determine whether it is stored in a locked room, locked cabinet, or area with restricted access. Check if the device is left unattended in unsecured areas such as open wards or unlocked procedure rooms.
    Affected if The device has weak or no physical access controls (e.g., stored in unlocked areas, accessible to unauthorized personnel)
  3. Examine the hard disk drive for tampering indicators
    Inspect the device for signs of HDD removal or tampering. Look for broken seals, unusual screw marks on the chassis, or evidence that the case has been opened. If possible, verify the HDD is properly seated and no unauthorized storage devices are connected.
    Affected if There are physical tampering indicators or unauthorized storage devices attached
  4. Review access logs for the device
    Check any available audit logs, access logs, or clinical documentation recording device usage. Look for unauthorized or unexplained access events, unusual login times, or access by unknown personnel.
    Affected if Access logs show unexplained or unauthorized access events
  5. Assess USB device connections
    Examine all USB ports on the device for connected peripherals. Document any USB devices that should not be present, such as unknown flash drives or USB hacking tools.
    Affected if Unknown or unauthorized USB devices are connected to the device
  6. Compare stored password hashes
    If you have administrative access, compare the current system passwords and account configurations against your documented baseline. Look for changes to local user accounts, administrator passwords, or clinical software access credentials.
    Affected if Passwords or account configurations have been changed without documented authorization

Your environment is affected if you have Boston Scientific Zoom Latitude Model 3120 devices with inadequate physical access controls or evidence of unauthorized physical access.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Physical access controls (restricted access to device storage, device locking) are the primary mitigations. Contact Boston Scientific for potential firmware updates or device replacement options, as this is a legacy device that may have limited vendor support.

Fix this in Zoom Latitude Pogrammer\/recorder\/monitor 3120 Firmware Scoped from the published advisory
  • Consultation8.0 h
  • Implementation12.0 h
  • Testing6.0 h
  • Review / QA4.0 h
30.0 hours of engineering $5,380
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $8,608.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2021-38400 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2021-38400 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data