CVE-2025-24836
Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.
NVD · uneditedWith a specially crafted Python script, an attacker could send continuous startMeasurement commands over an unencrypted Bluetooth connection to the affected device. This would prevent the device from connecting to a clinician's app to take patient readings and ostensibly flood it with requests, resulting in a denial-of-service condition.
Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.
dbcve analysis · moderate confidenceA denial-of-service vulnerability in a medical device allows an unauthenticated attacker to send continuous startMeasurement commands over unencrypted Bluetooth, preventing the device from connecting to a clinician's app and blocking patient readings. The attack exploits the lack of input validation and rate limiting on the Bluetooth interface.
Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.
CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.
From the vector- Attack vector
- Adjacent
- Complexity
- High
- Privileges
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- Low
- Integrity
- High
- Availability
- High
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:H/A:H
Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.
dbcve checksWork through these to decide whether this CVE applies to you.
-
Verify Bluetooth interface is enabled on the deviceAccess device management console or settings and check Bluetooth status, or run command: 'hciconfig' or check system Bluetooth service statusAffected if Bluetooth is enabled and exposed externally
-
Confirm Bluetooth encryption statusCheck Bluetooth security settings in device configuration. Look for 'encrypt' or 'secure connections' settings. On Linux: 'hciconfig hci0 auth' or check pairing/bonding statusAffected if Bluetooth is NOT using encryption (no BLE secure connections or pairing enabled)
-
Inspect rate limiting configuration for Bluetooth commandsCheck device configuration files or management interface for throttling/rate-limiting settings on Bluetooth commands, specifically for startMeasurement or similar measurement initiation commandsAffected if No rate limiting, throttling, or command throttling is configured for Bluetooth interfaces
-
Check authentication requirements for Bluetooth commandsReview Bluetooth security policies - verify if startMeasurement command requires authentication or pairing before execution. Check if unauthenticated access is permitted to Bluetooth interfacesAffected if Unauthenticated access to Bluetooth interface is allowed without pairing or authentication
-
Test for command injection or lack of input validationIf you have a test environment, send a series of rapid startMeasurement commands via Bluetooth and observe if the device accepts them without validation or blockingAffected if Device accepts excessive startMeasurement commands without validation or blocking
Your environment is affected if Bluetooth is enabled without encryption, no rate limiting exists on Bluetooth commands, and unauthenticated startMeasurement requests can be sent freely to block device connectivity.
Generated from the published advisory. Verify against your own configuration.
Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.
From vendor dataImplement Bluetooth encryption (e.g., BLE secure connections with pairing) and add command rate limiting/throttling on the device to reject excessive startMeasurement requests.
- Consultation8.0 h
- Implementation40.0 h
- Testing24.0 h
- Review / QA12.0 h
An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $23,296.
Scan for this in your stack
Free · runs locallyCheck whether your project pulls in CVE-2025-24836 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.
References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.
Primary sourcesPractitioner notes
ContributedPeer-ranked notes from engineers who’ve handled CVE-2025-24836 in production — separate from our analysis above.
The advisory tells you what broke. It rarely tells you what actually worked. If you’ve dealt with this one, that detail is what the next engineer is searching for.
- The version that genuinely resolved it — not the one the vendor claimed
- A config change or rule that shut the vector down
- A gotcha in the upgrade path that cost you an afternoon
No notes yet
Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.
A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.
- Verified mitigations, workarounds, and config changes
- Version or environment caveats, and links to real fixes
- No weaponised exploit code, or anything meant to cause harm