CWE-1263Weakness · CWE-1263

CVE-2025-4386

MEDIUM · 6.8 CVSS v3.1 Published 2026-05-07
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
73/100
Remediation priority · Elevated
No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Medtronic MyCareLink Patient Monitor has an internal serial interface, which allows an attacker with physical access to access a login prompt via a UART terminal.​

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

The Medtronic MyCareLink Patient Monitor contains an exposed internal serial (UART) interface that provides direct access to a login prompt. An attacker with physical access to the device can connect to this UART terminal and gain unauthenticated access to the system, potentially allowing further compromise of the patient monitor.

MitigationDisable or password-protect the UART interface in production firmware, and implement physical security controls to restrict access to the device internals.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Physical
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Confirm device model
    Locate the device identification label on the Medtronic MyCareLink Patient Monitor (usually on the bottom or rear of the unit) and verify it matches the affected product name
    Affected if The device is a Medtronic MyCareLink Patient Monitor and the UART interface is physically accessible
  2. Inspect device enclosure for UART access
    Open the device housing (check for screws, clips, or tamper-evident seals) and visually examine the main circuit board for exposed UART pins, test points, or a serial header connector typically arranged in a 3-4 pin configuration (RX, TX, GND, optionally VCC)
    Affected if UART pins or a serial header are exposed and connectable without requiring special tools or knowledge
  3. Test UART interface for authentication
    Connect a USB-to-TTL serial adapter to the exposed UART pins (connect adapter ground to board ground, RX to TX, TX to RX), power on the device, and use terminal software (e.g., PuTTY, screen) at common baud rates (115200, 57600, 9600) to observe the boot sequence and login prompt
    Affected if The UART terminal presents a login prompt or shell access without requiring any username or password
  4. Check for unauthenticated root access
    After connecting to the UART terminal, observe whether the system drops directly to a root shell or allows login as root/admin without credentials, or note any default credentials displayed in boot messages
    Affected if The UART interface provides root or administrator-level access without authentication
  5. Assess physical security controls
    Examine the device exterior for security features such as tamper-resistant screws, sealed enclosures, lockable casing, or security stickers that would deter or indicate unauthorized physical access to internals
    Affected if The device lacks physical security measures preventing easy access to internal UART interfaces

If physical inspection of a Medtronic MyCareLink Patient Monitor reveals an exposed UART interface that grants unauthenticated shell or root access, the device is affected by this vulnerability.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Disable or password-protect the UART interface in production firmware, and implement physical security controls to restrict access to the device internals.

Have this fixed Scoped from the published advisory
  • Consultation4.0 h
  • Implementation16.0 h
  • Testing8.0 h
  • Review / QA4.0 h
32.0 hours of engineering $5,600
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $8,960.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2025-4386 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2025-4386 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data