CVE-2025-4395
Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.
NVD · uneditedMedtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify system functionality. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.
dbcve analysis · high confidenceThe Medtronic MyCareLink Patient Monitor models 24950 and 24952 contain a hardcoded user account with an empty password. An attacker with physical access to the device can authenticate without credentials and modify system functionality.
Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.
CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.
From the vector- Attack vector
- Physical
- Complexity
- Low
- Privileges
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- High
- Integrity
- High
- Availability
- High
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.
dbcve checksWork through these to decide whether this CVE applies to you.
-
Identify the device model numberLocate the model number on the physical device label or in the device system information menu. Confirm it is either 24950 or 24952.Affected if The device model is 24950 or 24952.
-
Check if the hardcoded user account existsAttempt to access the device administration interface using the default credentials (typically admin or service account with no password). This may require physical access to the device or access to the device's management interface.Affected if You can authenticate without providing a password or using default credentials.
-
Verify device firmware versionAccess the device settings or system information screen to display the current firmware version. Compare against any firmware updates released by Medtronic before June 25, 2025.Affected if The firmware version is older than the corrected version released by Medtronic.
-
Inspect user account configurationIf you have administrative access to the device, navigate to the user management or account settings panel. Look for accounts with blank or empty password fields.Affected if User accounts with empty passwords are present in the device configuration.
You are affected if your device is model 24950 or 24952 and contains the hardcoded account with an empty password, which can be confirmed by attempting unauthenticated access or finding accounts with no password set.
Generated from the published advisory. Verify against your own configuration.
Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.
From vendor dataApply the vendor firmware update (released before June 25, 2025) or change the default account password to a strong, unique password through the device administration interface.
- Consultation4.0 h
- Implementation4.0 h
- Testing4.0 h
- Review / QA2.0 h
An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $3,968.
Scan for this in your stack
Free · runs locallyCheck whether your project pulls in CVE-2025-4395 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.
References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.
Primary sourcesPractitioner notes
ContributedPeer-ranked notes from engineers who’ve handled CVE-2025-4395 in production — separate from our analysis above.
The advisory tells you what broke. It rarely tells you what actually worked. If you’ve dealt with this one, that detail is what the next engineer is searching for.
- The version that genuinely resolved it — not the one the vendor claimed
- A config change or rule that shut the vector down
- A gotcha in the upgrade path that cost you an afternoon
No notes yet
Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.
A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.
- Verified mitigations, workarounds, and config changes
- Version or environment caveats, and links to real fixes
- No weaponised exploit code, or anything meant to cause harm
- No spam, self-promotion, credentials, or personal data