CVE-2025-4397
Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.
NVD · uneditedMedtronic MyCareLink Patient Monitor uses per-product credentials that are stored in a recoverable format. An attacker can use these credentials to modify encrypted drive data.
Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.
dbcve analysis · moderate confidenceThe Medtronic MyCareLink Patient Monitor stores per-product credentials in a recoverable format rather than using proper encryption or hashing. An attacker with access to these credentials can decrypt and modify data on the encrypted drive, compromising data integrity. The vulnerability stems from inadequate credential protection mechanisms in the device's storage subsystem.
Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.
CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.
From the vector- Attack vector
- Physical
- Complexity
- Low
- Privileges
- None
- User interaction
- None
- Scope
- Unchanged
- Confidentiality
- High
- Integrity
- High
- Availability
- High
CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.
dbcve checksWork through these to decide whether this CVE applies to you.
-
Identify device model and firmware versionLocate the Medtronic MyCareLink Patient Monitor and record the model number and firmware version from the device label or system information screen. If you have administrative access, use the device's built-in diagnostic or settings menu to retrieve the firmware version.Affected if The firmware version is any version prior to the vendor-supplied patch that addresses inadequate credential protection.
-
Verify credential storage accessDetermine if you have administrative or service-level access to the device filesystem or configuration backend. This may require physical access to the device or authenticated remote access to its management interface.Affected if You can access the credential storage subsystem - the vulnerability is present if credentials are accessible in any recoverable format.
-
Examine stored credential formatIf you have filesystem access, locate the credential storage directory (commonly found in device configuration or security-related directories). Inspect the credential files using a hex editor or text viewer to determine if the credentials appear as plain text, Base64-encoded, or otherwise reversibly encoded data.Affected if Credentials are stored in any format that can be reversed to plaintext without requiring a cryptographic key derivation process.
-
Check for proper encryption implementationExamine whether the credential storage uses proper cryptographic encryption (such as AES with secure key management) or relies on simple encoding or obfuscation. Look for initialization vectors, salted hashes, or key derivation functions in the storage mechanism.Affected if The credential storage lacks proper encryption and uses only encoding, obfuscation, or reversible transformation methods.
-
Review access controls on credential storageAudit file permissions or access control lists on credential storage locations. Determine if the storage is protected by system-level encryption or if credentials are stored in a location accessible to less-privileged processes.Affected if Credentials are stored in a recoverable format and the storage location has weak access controls, allowing potential extraction.
A user is affected if the Medtronic MyCareLink device stores credentials in a reversible (non-encrypted) format that could allow an attacker with physical or administrative access to extract and decrypt them.
Generated from the published advisory. Verify against your own configuration.
Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.
From vendor dataApply vendor-supplied patches or firmware updates from Medtronic that implement proper credential encryption. Until the patch is available, restrict physical and network access to the device to limit credential exposure.
- Consultation8.0 h
- Implementation4.0 h
- Testing12.0 h
- Review / QA6.0 h
An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $8,320.
Scan for this in your stack
Free · runs locallyCheck whether your project pulls in CVE-2025-4397 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.
References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.
Primary sourcesPractitioner notes
ContributedPeer-ranked notes from engineers who’ve handled CVE-2025-4397 in production — separate from our analysis above.
The advisory tells you what broke. It rarely tells you what actually worked. If you’ve dealt with this one, that detail is what the next engineer is searching for.
- The version that genuinely resolved it — not the one the vendor claimed
- A config change or rule that shut the vector down
- A gotcha in the upgrade path that cost you an afternoon
No notes yet
Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.
A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.
- Verified mitigations, workarounds, and config changes
- Version or environment caveats, and links to real fixes
- No weaponised exploit code, or anything meant to cause harm