Xper Information Management Physiomonitoring 5Application · Philips

CVE-2013-2808

HIGH · 9.3 CVSS v2.0 Published 2013-10-05
Fix available
A fix is available. Upgrade to after 1.5.4.053 or later.
See remediation →
100/100
Remediation priority · Urgent
Remotely reachable No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Heap-based buffer overflow in Xper in Philips Xper Information Management Physiomonitoring 5 components, Xper Information Management Vascular Monitoring 5 components, and Xper Information Management servers and workstations for Flex Cardio products before XperConnect 1.5.4.053 SP2 allows remote attackers to execute arbitrary code via a crafted HTTP request to the Connect broker on TCP port 6000.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

Heap-based buffer overflow vulnerability in Philips Xper Information Management system components (Physiomonitoring 5, Vascular Monitoring 5, and Flex Cardio products). Remote attackers can execute arbitrary code via crafted HTTP requests to the Connect broker service listening on TCP port 6000.

MitigationUpgrade to XperConnect 1.5.4.053 SP2 or later. Until patched, isolate the Connect broker (port 6000) from untrusted networks using firewalls or network segmentation to reduce attack surface.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Xper Information Management Physiomonitoring 5Application
Affected:all versions
XperconnectApplication
Affected:<= 1.5.4.053
Xper Information Management Vascular Monitoring 5Application
Affected:all versions
Xper Flex CardioHardware / appliance
Affected:all versions

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Network
Complexity
M
Authentication
None
Confidentiality
Complete
Integrity
Complete
Availability
Complete

AV:N/AC:M/Au:N/C:C/I:C/A:C

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify installed Philips Xper components
    Check system inventory or installed programs for Philips Xper Information Management Physiomonitoring 5, Vascular Monitoring 5, Flex Cardio, or XperConnect
    Affected if Any of these products are present and the Connect broker service is accessible on port 6000
  2. Determine XperConnect version
    Locate the XperConnect installation directory and check the version information (typically in application metadata or version file)
    Affected if XperConnect version is 1.5.4.053 or lower
  3. Verify Connect broker service status
    Check if the Connect broker service is running and listening on TCP port 6000 using netstat or equivalent network scanning tool
    Affected if Port 6000 is open and accepting connections on the affected system
  4. Confirm product is in affected family
    Confirm the system runs Philips Xper Information Management Physiomonitoring 5, Vascular Monitoring 5, or Flex Cardio regardless of version, or XperConnect version 1.5.4.053 or below
    Affected if The product is any version of Physiomonitoring 5, Vascular Monitoring 5, or Flex Cardio, or XperConnect version 1.5.4.053 or below, AND the Connect broker on port 6000 is network-accessible

The system is affected if it runs any Philips Xper Information Management component (Physiomonitoring 5, Vascular Monitoring 5, Flex Cardio all versions, or XperConnect version 1.5.4.053 or lower) AND has the Connect broker service on TCP port 6000 exposed and running.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

dbcve · scoped
Upgrade available Upgrade to a release after 1.5.4.053
Interim mitigation

Upgrade to XperConnect 1.5.4.053 SP2 or later. Until patched, isolate the Connect broker (port 6000) from untrusted networks using firewalls or network segmentation to reduce attack surface.

Recommended fix High confidence

XperConnect 1.5.4.053 SP2

  1. Upgrade XperConnect to version 1.5.4.053 SP2 or later to remediate the heap-based buffer overflow vulnerability
  2. Apply the upgrade to all Xper Information Management servers and workstations running Flex Cardio products
  3. After upgrading, verify the Connect broker on TCP port 6000 is running the patched version

Generated from the published advisory — verify against the referenced sources before acting.

Fix this in Xper Information Management Physiomonitoring 5 Scoped from the published advisory
  • Consultation8.0 h
  • Implementation6.0 h
  • Testing8.0 h
  • Review / QA4.0 h
26.0 hours of engineering $4,600
Get the upgrade done

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $7,360.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2013-2808 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2013-2808 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data