Infinity Explorer C700 FirmwareOperating system · Draeger

CVE-2019-25718

HIGH · 7.8 CVSS v3.1 Published 2026-06-01
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
80/100
Remediation priority · High
Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Dräger Infinity Explorer C700 contains a privilege escalation vulnerability that allows attackers to break out of kiosk mode and access the underlying operating system through a specific dialog interaction. Attackers can exploit this kiosk escape to take control of the operating system and cause the device to display incorrect or no information from the connected Delta Family patient monitor.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

A kiosk mode escape vulnerability in Dräger Infinity Explorer C700 allows attackers to break out of the restricted interface through a specific dialog interaction, gaining access to the underlying Windows operating system. Once escaped, attackers can take full control of the OS and interfere with the connection to Delta Family patient monitors, potentially causing incorrect or absent patient data display.

MitigationApply vendor-supplied patches for Infinity Explorer C700, implement application whitelisting and kiosk lockdown controls, restrict physical access to the device, and ensure network segmentation. Coordinate with Dräger for validated remediation given the medical device context.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Infinity Explorer C700 FirmwareOperating system
Affected:all versions

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Local
Complexity
Low
Privileges
Low
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify device model
    Locate the medical device and confirm it is a Dräger Infinity Explorer C700 monitor station. Check device labeling, system information, or clinical documentation for the model number.
    Affected if The device is confirmed to be a Dräger Infinity Explorer C700 and runs the Dräger Infinity Explorer software.
  2. Confirm kiosk mode configuration
    Access the device settings or configuration panel to verify whether kiosk mode (restricted user mode) is enabled. Look for settings that lock the user interface to a specific application or prevent access to the Windows desktop.
    Affected if Kiosk mode is enabled on the device, which is the required condition for the kiosk escape vulnerability to apply.
  3. Inspect for unauthorized Windows access
    Check Windows event logs, user accounts, and recent session activity on the device for signs of desktop access that occurred outside the normal kiosk-mode application flow. Look for new user accounts, modified permissions, or unexpected processes.
    Affected if Evidence exists of unauthorized access to the Windows desktop or operating system beyond the intended kiosk-mode application.
  4. Review Delta Family monitor data integrity
    Compare the patient monitoring data displayed on the C700 with the actual output from connected Delta Family monitors. Check for discrepancies, missing data, or signs of manipulation in the data stream or display.
    Affected if Patient monitoring data from connected Delta Family monitors shows unexpected discrepancies or is missing, suggesting possible manipulation after a kiosk escape.
  5. Audit physical and peripheral access
    Inspect the device for unauthorized USB devices, external storage, or physical tampering. Review access logs if available for the physical location where the device is deployed.
    Affected if Physical security controls are inadequate or unauthorized peripheral devices are connected, increasing the likelihood of kiosk escape exploitation.

A user is affected if they operate a Dräger Infinity Explorer C700 device with kiosk mode enabled and there is evidence of unauthorized Windows desktop access or manipulation of Delta Family monitor data.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Apply vendor-supplied patches for Infinity Explorer C700, implement application whitelisting and kiosk lockdown controls, restrict physical access to the device, and ensure network segmentation. Coordinate with Dräger for validated remediation given the medical device context.

Fix this in Infinity Explorer C700 Firmware Scoped from the published advisory
  • Consultation8.0 h
  • Implementation16.0 h
  • Testing16.0 h
  • Review / QA8.0 h
48.0 hours of engineering $8,320
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $13,312.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2019-25718 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2019-25718 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data