M1000 Multipara Patient Monitor FirmwareOperating system · Niscomed

CVE-2020-15482

HIGH · 7.8 CVSS v3.1 Published 2020-08-26
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
80/100
Remediation priority · High
Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
An issue was discovered on Nescomed Multipara Monitor M1000 devices. The device enables an unencrypted TELNET service by default, with a blank password for the admin account. This allows an attacker to gain root access to the device over the local network.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

The Nescomed Multipara Monitor M1000 ships with an unencrypted TELNET service enabled by default, utilizing a blank password for the admin account. This configuration allows any local network attacker to connect anonymously and gain root shell access to the device without authentication.

MitigationDisable the TELNET service if not operationally required; if TELNET is necessary, implement strong authentication and network access controls. Consider network segmentation to isolate medical devices from untrusted networks.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
M1000 Multipara Patient Monitor FirmwareOperating system
Affected:all versions

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Local
Complexity
Low
Privileges
Low
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

CVSS:3.1/AV:L/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Verify TELNET port accessibility
    Scan the target device for open port 23 (TELNET) using nmap or similar network scanner: `nmap -p 23 <device_ip>`
    Affected if Port 23 is open and reachable from your network segment
  2. Test for unauthenticated TELNET access
    Attempt an anonymous TELNET connection to the device: `telnet <device_ip>`. Observe whether a shell prompt is provided without requiring credentials.
    Affected if Connection succeeds and provides shell access without any authentication prompt
  3. Check admin account authentication
    If prompted for credentials during TELNET attempt, try the admin account with blank password: login as 'admin' with no password or empty password field.
    Affected if Access is granted with blank admin password, indicating the default vulnerable configuration is present
  4. Assess network exposure
    Review network segmentation and access control lists to determine if the device TELNET service is reachable from untrusted network segments or the broader local network.
    Affected if Device is accessible from network segments outside of isolated management VLANs or from untrusted systems

Your environment is affected if the Nescomed Multipara Monitor M1000 has TELNET port 23 accessible and allows unauthenticated root shell access, which is the default vulnerable configuration for all firmware versions.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Disable the TELNET service if not operationally required; if TELNET is necessary, implement strong authentication and network access controls. Consider network segmentation to isolate medical devices from untrusted networks.

Fix this in M1000 Multipara Patient Monitor Firmware Scoped from the published advisory
  • Consultation4.0 h
  • Implementation2.0 h
  • Testing3.0 h
  • Review / QA3.0 h
12.0 hours of engineering $2,150
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $3,440.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2020-15482 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2020-15482 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data