Mycarelink Smart Model 25000 FirmwareOperating system · Medtronic

CVE-2020-27252

HIGH · 8.1 CVSS v3.1 Published 2020-12-14
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
90/100
Remediation priority · Urgent
Remotely reachable No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
Medtronic MyCareLink Smart 25000 is vulnerable to a race condition in the MCL Smart Patient Reader software update system, which allows unsigned firmware to be uploaded and executed on the Patient Reader. If exploited, an attacker could remotely execute code on the MCL Smart Patient Reader device, leading to control of the device.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

CVE-2020-27252 is a race condition vulnerability in the MCL Smart Patient Reader software update system that allows unsigned firmware to be uploaded and executed, enabling remote code execution and full control of the Medtronic MyCareLink Smart 25000 device.

MitigationApply vendor-supplied patches when available and restrict network accessibility of the device until the update system is properly secured with cryptographic code signing verification.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Mycarelink Smart Model 25000 FirmwareOperating system
Affected:all versions

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Network
Complexity
High
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify device model
    Locate the Medtronic MyCareLink Smart device and verify the model number is 25000. Check device packaging, physical label, or management interface for 'Model 25000' or 'MyCareLink Smart' branding.
    Affected if Device is a Medtronic MyCareLink Smart Model 25000
  2. Check firmware version
    Access the device management interface or use the manufacturer's diagnostic tool to retrieve the current firmware version. Compare against the affected range.
    Affected if Any firmware version is present on a Model 25000 device (all versions affected)
  3. Verify network accessibility of update service
    Check if the device is network-connected and if ports or services related to the software update mechanism are exposed. Use network scanning to identify open ports on the device IP address.
    Affected if Device update service is network-accessible from untrusted networks
  4. Inspect update mechanism configuration
    Examine the device or associated software configuration for the software update system settings. Look for code signing verification flags or update authentication parameters.
    Affected if Code signing verification for firmware updates is disabled or not enforced

You are affected if you have a Medtronic MyCareLink Smart Model 25000 device with any firmware version and the software update system is network-accessible or lacks code signing verification.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Apply vendor-supplied patches when available and restrict network accessibility of the device until the update system is properly secured with cryptographic code signing verification.

Fix this in Mycarelink Smart Model 25000 Firmware Scoped from the published advisory
  • Consultation8.0 h
  • Implementation40.0 h
  • Testing16.0 h
  • Review / QA8.0 h
72.0 hours of engineering $12,640
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $20,224.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2020-27252 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2020-27252 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data