Anydana A FirmwareOperating system · Sooil

CVE-2020-27264

HIGH · 8.8 CVSS v3.1 Published 2021-01-19
Fix available
A fix is available. Upgrade to 3.0 or later.
See remediation →
93/100
Remediation priority · Urgent
No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications use deterministic keys, which allows unauthenticated, physically proximate attackers to brute-force the keys via Bluetooth Low Energy.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

The Diabecare RS insulin pump and its companion AnyDana-i and AnyDana-A mobile applications use deterministic (static) keys for BLE authentication. Attackers within Bluetooth range can perform offline brute-force attacks against these predictable keys to inject unauthorized commands to the pump, potentially manipulating insulin delivery.

MitigationReplace deterministic keys with cryptographically random, session-based keys or implement proper BLE authentication with secure key exchange. Coordinate with the vendor for firmware and application updates, then validate the fix in an isolated environment before deploying to clinical settings.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Anydana A FirmwareOperating system
Affected:< 3.0
Anydana I FirmwareOperating system
Affected:< 3.0
Diabecare Rs FirmwareOperating system
Affected:< 3.0

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Adjacent
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
High
Availability
High

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Verify BLE is enabled on the pump
    Access the Diabecare RS pump settings or status menu and confirm the Bluetooth Low Energy radio is active. Alternatively, check if the pump is visible to BLE scanning tools or paired with the companion app.
    Affected if BLE is currently enabled - the vulnerability requires an active BLE connection to exploit.
  2. Locate the firmware version in pump settings
    Navigate to the device information or system settings screen on the Diabecare RS insulin pump. Look for a 'Firmware Version', 'Software Version', or 'Device Info' entry that displays a numeric version number.
    Affected if The displayed firmware version is below 3.0 (for example, 2.x.x or 1.x.x).
  3. Check AnyDana-A app version
    Open the AnyDana-A mobile application on the paired Android/iOS device, then go to Settings > About or App Info to view the installed application version.
    Affected if The companion mobile app is running and the device firmware is version 3.0 or older.
  4. Check AnyDana-i app version
    Open the AnyDana-i mobile application, navigate to Settings > About, and record the application version displayed.
    Affected if The companion mobile app is running and the device firmware is version 3.0 or older.

Your environment is affected if BLE is enabled on the pump AND the firmware version is less than 3.0 for any of the Anydana-A, Anydana-I, or Diabecare RS devices.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Upgrade available Upgrade to 3.0 or later
Fixed in 3.0
Interim mitigation

Replace deterministic keys with cryptographically random, session-based keys or implement proper BLE authentication with secure key exchange. Coordinate with the vendor for firmware and application updates, then validate the fix in an isolated environment before deploying to clinical settings.

Fix this in Anydana A Firmware Scoped from the published advisory
  • Consultation12.0 h
  • Testing16.0 h
  • Review / QA8.0 h
36.0 hours of engineering $6,240
Get the upgrade done

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $9,984.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2020-27264 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2020-27264 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data