Mycarelink Monitor 24950 FirmwareOperating system · Medtronic

CVE-2019-6540

MEDIUM · 6.5 CVSS v3.1 Published 2019-03-26
Mitigation only
No fix yet — a mitigation exists. There is no fixed release. A documented workaround reduces exposure in the meantime.
See remediation →
70/100
Remediation priority · Elevated
No privileges Zero-click

Official description Straight from the sourceThe vendor's or NVD's own wording, published unedited. Authoritative, but often terse — it says what broke, rarely what to do.

NVD · unedited
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.

Technical summary Written by usOur analysis, written from the advisory, the CVSS vector and the affected-version data. It adds context the advisory leaves out, and never invents facts that are not in the source.

dbcve analysis · high confidence

The Conexus telemetry protocol used in Medtronic cardiac devices (MyCareLink Monitor, CareLink Monitor/Programmer, and various ICD/CRT-D devices) lacks encryption implementation. An attacker with adjacent short-range wireless access can passively intercept unencrypted communications containing sensitive patient data and device telemetry.

MitigationImplement compensating controls including physical security around devices to limit proximity access, network segmentation to isolate medical devices, and monitor for unauthorized wireless listeners. Apply vendor firmware updates when available. Since the protocol itself lacks encryption, replacement or upgrade of affected devices to versions with encrypted telemetry may ultimately be required.

Verify against the referenced sources before acting — the references below are authoritative for this CVE, this summary is not.

Affected products & versions What the vendor confirmedThe version ranges the vendor confirmed as vulnerable. If your version sits inside a range here, treat yourself as exposed until you have upgraded.

NVD · CPE data
Mycarelink Monitor 24950 FirmwareOperating system
Affected:all versions
Mycarelink Monitor 24952 FirmwareOperating system
Affected:all versions
Carelink Monitor 2490c FirmwareOperating system
Affected:all versions
Carelink 2090 FirmwareOperating system
Affected:all versions
Amplia Crt D FirmwareOperating system
Affected:all versions
Claria Crt D FirmwareOperating system
Affected:all versions
Compia Crt D FirmwareOperating system
Affected:all versions
Concerto Crt D FirmwareOperating system
Affected:all versions

CVSS breakdown How the score is builtThe industry scoring standard. It rates how the flaw is reached, what it takes to exploit, and what an attacker gains — the score is derived from those, not the other way round.

From the vector
Attack vector
Adjacent
Complexity
Low
Privileges
None
User interaction
None
Scope
Unchanged
Confidentiality
High
Integrity
None
Availability
None

CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N

Am I affected? How to checkSteps we derive from the advisory and the affected-version data, so you can decide whether this CVE reaches your setup. They are a guide, not a scan — your own configuration is the authority.

dbcve checks

Work through these to decide whether this CVE applies to you.

  1. Identify Medtronic cardiac device model and firmware version
    Locate the device inventory or physically inspect the Medtronic cardiac device (MyCareLink Monitor, CareLink Monitor, or ICD/CRT-D device). Record the model number (e.g., 24950, 24952, 2490c, 2090, Amplia, Claria, Compia, Concerto) and firmware version. This information is typically found on the device label or in the device management software.
    Affected if The device model matches one of the affected products (24950, 24952, 2490c, 2090, Amplia, Claria, Compia, Concerto) and it is running any firmware version.
  2. Verify Conexus telemetry protocol is active
    Check the device configuration settings or clinical system logs for the Conexus telemetry interface. In the Medtronic device programmer or monitor settings, look for telemetry status, wireless communication status, or remote monitoring settings. Confirm that the Conexus protocol is enabled for wireless communication.
    Affected if Conexus telemetry is enabled and actively transmitting data via short-range wireless.
  3. Confirm absence of encryption on telemetry traffic
    Using a wireless spectrum analyzer or network monitoring tool capable of receiving the device's short-range wireless signals (typically in the MICS band used for medical implant communication), capture and inspect the telemetry traffic. Look for unencrypted plaintext transmission of patient data or device telemetry. Since the protocol lacks encryption by design, traffic will be readable in cleartext.
    Affected if Captured telemetry traffic is transmitted in unencrypted, plaintext format that can be read by an adjacent wireless listener.
  4. Assess physical proximity controls around the device
    Inspect the physical environment where the Medtronic device or its bedside monitor is located. Document the accessibility of the device to unauthorized persons, the presence of physical barriers, and the proximity of the device to public or uncontrolled areas.
    Affected if The device is located in an area where an attacker with short-range wireless access (within a few meters) could feasibly intercept the telemetry signals.

A defender is affected if they have deployed any of the listed Medtronic devices (MyCareLink Monitor, CareLink Monitor, or CRT-D/ICD devices) with Conexus telemetry enabled, as all firmware versions lack encryption and the unencrypted wireless communications can be intercepted by an attacker within range.

Generated from the published advisory. Verify against your own configuration.

Check your environment

Paste your version and any relevant configuration and it will be compared against the affected criteria above. Do not include secrets or credentials.

AI-assisted, checked against the advisory. Informational, not a guarantee.

Remediation Closing itWhat it takes to close this. Where a vendor fix exists we point at it; where none exists we say so plainly, and can build one. Effort estimates are scoped from the advisory, not from your codebase.

From vendor data
Mitigation available No clean upgrade yet — mitigate in the meantime
Mitigation

Implement compensating controls including physical security around devices to limit proximity access, network segmentation to isolate medical devices, and monitor for unauthorized wireless listeners. Apply vendor firmware updates when available. Since the protocol itself lacks encryption, replacement or upgrade of affected devices to versions with encrypted telemetry may ultimately be required.

Fix this in Mycarelink Monitor 24950 Firmware Scoped from the published advisory
  • Consultation16.0 h
  • Implementation24.0 h
  • Testing20.0 h
  • Review / QA12.0 h
72.0 hours of engineering $12,680
Get help mitigating

An estimate, not a bill — we confirm scope with you before any work starts. Need it this week? Rush from $20,288.

Scan for this in your stack

Free · runs locally
dbcve dependency scanner

Check whether your project pulls in CVE-2019-6540 — or any other known-vulnerable package — straight from your lock files. Free and open source; it runs locally and uploads nothing.

References Go to the primary sourcePrimary sources — vendor advisories, patches and trackers. Where our summary and a reference disagree, the reference wins.

Primary sources

Practitioner notes

Contributed

Peer-ranked notes from engineers who’ve handled CVE-2019-6540 in production — separate from our analysis above.

No notes yet

Be the first to add a field note for this CVE — a mitigation you’ve verified, a version caveat, or a link to a working fix. Sign in above to contribute.

What this is

A place for practitioners to share what actually worked: a mitigation you’ve tested, a configuration change, a version- or environment-specific caveat, or a link to a verified patch. The most useful notes rise to the top as peers upvote them, so the signal stays high.

What belongs here
  • Verified mitigations, workarounds, and config changes
  • Version or environment caveats, and links to real fixes
  • No weaponised exploit code, or anything meant to cause harm
  • No spam, self-promotion, credentials, or personal data